Troubleshooting - No staining
Immunohistochemistry (IHC) troubleshooting for no staining, high background and non-specific staining.
Problems | Possible reasons | Solutions |
---|---|---|
No staining | Wrong primary antibody | Choose an antigen-specific antibody |
Inappropriate concentration (usually too low) of primary or secondary antibody | Optimize the concentration | |
The activities of primary/secondary antibody lost. | Check the antibody storage condition and run a positive control | |
Inappropriate pH of antibody diluents | Make sure antibody works properly | |
Insufficient deparaffinization | Deparaffinize sections longer | |
Insufficient permeabilization | Add an appropriate permeabilizing agent | |
Insufficient antigen retrieval | Follow the antigen retrieval protocol | |
The secondary antibody was not stored in the dark | Biotinylated secondary antibody should be stored in the dark. |
Learn more about:
- Immunohistochemistry Troubleshooting-High Background
- Immunohistochemistry Troubleshooting-No specific staining
- Immunohistochemistry (IHC) principle
- Immunohistochemistry (IHC) protocol
- Immunohistochemistry (IHC) staining
- Immunohistochemistry (IHC) antibody
- Immunohistochemistry (IHC) application
- Immunohistochemistry (IHC) Tips